Polydisperse Particle Standards - Nominal Size Range:
|
10
- 100 microns |
||||||
|
Table of Results |
Percentile | |||||
|
Method |
Laboratory |
10 | 25 | 50 | 75 | 90 |
| X | 25.10 | 32.30 |
41.70 |
53.70 |
62.40 |
|
| L | 25.00 |
31.50 |
41.00 |
51.70 |
61.70 |
|
| E | 24.50 |
32.50 |
41.80 |
51.80 |
62.40 |
|
| Mean | 24.87 |
32.10 |
41.50 |
52.40 |
62.17 |
|
| SD | 0.26 |
0.43 |
0.36 |
0.92 |
0.33 |
|
| C | 25.63 |
32.11 |
42.46 |
54.18 |
64.59 |
|
| M | 25.39 |
31.34 |
41.59 |
53.65 |
64.34 |
|
| J | 25.48 |
32.69 |
43.43 |
55.09 |
66.12 |
|
| D | 24.92 |
30.29 |
39.85 |
51.52 |
61.69 |
|
| X | 24.81 |
30.70 |
40.50 |
51.77 |
62.11 |
|
| U | 24.69 |
30.03 |
39.36 |
50.62 |
60.84 |
|
| W | 23.98 |
29.59 |
39.20 |
50.73 |
61.29 |
|
| Mean | 24.99 |
30.96 |
40.91 |
52.51 |
63.00 |
|
| SD* | 0.53 | 1.05 |
1.50 |
1.65 | 1.86 | |
F |
26.80 |
33.90 |
42.90 |
54.20 |
64.50 |
|
| Q | 25.30 |
31.00 |
39.80 |
52.00 |
62.90 |
|
X1 |
25.50 |
30.70 |
39.60 |
51.80 |
62.10 |
|
X2 |
28.20 |
34.90 |
44.40 |
53.90 |
64.10 |
|
Mean |
26.45 |
32.63 |
41.68 |
52.98 |
63.40 |
|
SD* |
1.16 |
1.80 |
2.05 |
1.08 |
0.95 |
|
| Final Mean size | x(n=14) |
25.37 |
31.68 |
41.26 |
52.62 |
62.93 |
| *SD (population) | xs n |
1.00 |
1.44 |
1.57 |
1.39 |
1.48 |
| SD (sample) | xs n-1 |
1.04 |
1.50 |
1.63 |
1.44 |
1.54 |
| Uncertainty | 95% confidence |
2.08 |
3.00 |
3.26 |
2.88 |
3.08 |
|
Graphical Review |
Graphical Review |
|
|
|
The Leading European Laboratories in this section were: |
|
| Bradford University | UK |
| Coulter Electronics | UK |
| Delft TU | Holland |
| IFTS | France |
| INRS | France |
| Loughborough University | UK |
| Seescan | UK |
| Solvay et Cie | Belgium |
| Whitehouse Scientific | UK |