NIST Traceable Sieve Standards
- Narrow Distribution Glass Microspheres

 

 

Unique method of calibrating sieves from 20 - 3350mm

Certified using Electroform sieves NIST & NPL traceable  

Accurate to 1mm for the smaller aperture sieves

High speed calibration - complete in about 2 mins

Single shot bottles remove operator sampling errors

Aperture sizes independent of sieve shaking method

Analyses over 80% of the sieves surface

Sieves don't need to be sent away for calibration

 

 

WHICH STANDARD SHOULD I USE?

The aperture size of every sieve manufactured from 20 to 3350 microns can be certified to NIST traceability.
Our standards are named after the 30 most popular sieves (R40/3) but can also be used for the 30 intermediate sizes.
If you are unsure of the standard you should be using just check the table below, the ‘primary’ sieve size is highlighted in bold, the sieve on either side can also be calibrated using the same Sieve Standard.

 

For Sieve
Size  
(
mm)
Sieve
Diameter
(mm)
Mesh
#
Calibration Range Wt. per Bottle (g) Catalogue 
Number
Price Per 5 bottle Set
£ €* $*
20 200 635 18.8 - 23.7 0.8 SS391 96.50 116 154
25 200 500 21.7 - 30.2 0.8 SS392 88.50 106 142
32 200 450 27.8 - 34.1 1.0 SS393 82.00 98 131
36, 38, 40 200 400 33.5 - 41.6 1.0 SS394 82.00 98 131
   45, 50 200 325 42.0 - 50.8 1.0 SS395 78.50 94 126
   53, 56 200 270 48.4 - 59.5 1.0 SS396 78.50 94 126
 63 200 230 56.6 - 70.4 1.0 SS397 77.50 93 124
71, 75, 80 200 200 67.1 - 82.8 1.0 SS398 76.50 92 122
90 200 170 78.8 - 97.6 1.0 SS399 76.50 92 122
100, 106, 112 200 140 91.4 - 117 1.0 SS400 76.50 92 122
125 200 120 112 - 139 1.0 SS401 76.50 92 122
140, 150, 160 200 100 134 - 167 1.5 SS402 76.50 92 122
180 200 80 161 - 199 1.5 SS403 76.50 92 122
200, 212, 224 200 70 191 - 237 1.5 SS404 76.50 92 122
   250, 280 200 60 226 - 281 2.5 SS405 72.00 86 115
   300, 315 200 50 270 - 333 2.5 SS406 72.00 86 115
355 200 45 322 - 398 2.5 SS407 72.00 86 115
400, 425, 450 200 40 377 - 470 2.5 SS408 72.00 86 115
500 200 35 440 - 557 2.5 SS409 72.00 86 115
560, 600, 630 200 30 526 - 657 2.5 SS410 72.00 86 115
710 200 25 658 - 809 2.5 SS411 72.00 86 115
800, 850, 900 200 20 774 - 951 2.5 SS412 72.00 86 115
1000 200 18 910 - 1106 7.0 SS413 92.00 110 147
1120, 1180, 1250 200 16 1091-  1335 10.0 SS414 110.50 133 177
     1400, 1550 200 14 1292 - 1609 15.0 SS415 125.00 150 200
1600, 1700, 1800 200 12 1515 - 1866 15.0 SS416 125.00 150 200
2000 200 10 1836 - 2236 20.0 SS417 128.50 155 206
2240, 2360, 2500 200 8 2148 - 2661 20.0 SS418 128.50 155 206
     2800, 3150 200 7 2555 - 3232 25.0 SS419 155.00 186 248
     3350, 3550 200 6 3072 - 3783 25.0 SS420 155.00 186 248

  * Currency conversions are provided as a guide only and are therefore subject to slight fluctuation

 

STANDARDS FOR SMALL DIAMETER SIEVES
Standards for 40mm and 100mm sieve are available on request.

 

RECOMMENDED NUMBER OF BOTTLES TO USE
FOR LARGER DIAMETER SIEVES
 

200mm Diameter Sieve
1 bottle per test
 

300mm Diameter Sieve
2 bottles per test
 

450mm Diameter Sieve
5 bottles per test 
 

1000mm Diameter Sieve
10 bottles* per test
  
*For a 1000mm sieve 10 bottles equals approximately 30% coverage of the sieve surface as apposed to our usual 80%.  It is  therefore recommended 3 tests are performed by the operator to ensure good repeatability of the test.  In order to cover 80% of the sieve surface 25 bottles would be required.
 

2000mm Diameter Sieve
20 bottles* per test
 
*For a 2000mm sieve 20 bottles equals approximately 30% coverage of the sieve surface as apposed to our usual 80%.  It is therefore recommended 3 tests are performed by the operator to ensure good repeatability of the test.  In order to cover 80% of the sieve surface 100 bottles would be required.


 

 

 

  

 

INSTRUCTIONAL ANIMATION
on a 30 second loop

INSTRUCTIONS

SIEVE CALIBRATION BY THE GLASS MICROSPHERE METHOD

 
Instructions

Place the 200mm or 8 inch sieve to be calibrated with the collecting pan on a 0.01g resolution balance and tare. Select the appropriate calibration standard for the sieve and a pour on a single-shot bottle. Record the initial weight of the microspheres. Shake the standard over the surface of the sieve using one of the generic methods shown above until the end-point is reached - see below. When complete, tap the frame a few times to dislodge near mesh beads and empty the undersize fraction from the pan into a collecting vessel. (These microspheres can be kept for future analysis by microscope if the maximum aperture size of the sieve needs to be determined). Reassemble the sieve and pan and tap a few more times by hand. If beads still fall through the mesh, the shaking time needs to be increased because the end-point has not been reached. Empty the pan again if necessary. Without resetting the tare on the balance, re-weigh the sieve and pan together with the retained microspheres. Record the weight. From the retained weight, calculate the percentage of microspheres passing the sieve and use the calibration graph supplied with the test certificate to determine the mean aperture size.

 
Sieve shaking methods

Use a vigorous swirling action to disperse the standard over the sieve surface. 2–3 cycles per second for 1 minute is recommended.

Shaking times may vary from 1 to 3 minutes depending on the sieve shaker. Empty and check the pan each minute to determine the end -point.

A vacuum of 2000–2200Pa for 3 minutes is adequate for most sizes above 30 um. The end- point is when the retained fraction is constant.

Run time typically 1 minute. An amplitude of 30 is sufficient to fluidize most standards but increase if necessary.

 
Mean Aperture Calculation

1. Calculate the percentage of the microspheres passing.
2. Read off the mean aperture size from the calibration graph or use the Whitehouse Scientific Sieve Aperture Size Calculator Software.

Notes:
(a) For sieves below 100um a 5% difference in weight passing usually only corresponds to a 1um difference in aperture size, which makes this method one of the most accurate ways of calibrating a sieve.
(b) To clean the sieve, lightly brush the underside with a soft bristle bush or use an ultrasonic bath. Never use a wire brush or sharp object to remove trapped beads.
(c) For 300 and 450mm sieves, use 2 to 5 bottles (see library).


A calibration graph from a 63
mm sieve standard

 

Example Certificate

Each standard is supplied with a certificate of analysis,
to see a PDF example of one please click below:

 

 

Tolerances and Recommended Aperture Sizes in the Calibration of 200mm Diameter Test Sieves by Microscopy 
- from BS 410-1:2000, ISO 3310-1:2000

Sieve Size (mm) Tolerance on Aperture Size (mm) Maximum Size for Any Aperture (mm)

Recommended Count for Microscopy

20 17.7 - 22.3 34 2 x 300
25 22.5 - 27.5 41 2 x 300
32 29.3 - 34.7 51 2 x 300
38 35.1 - 40.9 58 2 x 300
45 41.9 - 48.1 67 2 x 250
53 49.6 - 56.4 77 2 x 250
63 59.3 - 66.7 89 2 x 250
75 70.9 - 79.1  104 2 x 250
90 85.4 - 94.6 122 2 x 200
106 100.8 - 111.2 141 2 x 200
125 119.2 - 130.8 163 2 x 200
150 143.4 - 156.6 193 2 x 200
180 172.4 - 187.6 227 2 x 200
212 203.3 - 220.7 264 2 x 160
250 240.1 - 259.9 308 2 x 160
300 288 - 312 365 2 x 160
355 342 - 368 427 2 x 160
425 409 - 441 506 2 x 120
500 482 - 518 589 2 x 120
600 579 - 621 701 2 x 100
710 685 - 735 822 2 x 100
850 821 - 879 977 2 x 80
1000 970 - 1030 1140 2 x 80
1180 1140 - 1220 1340 2 x 80
1400 1350 - 1450 1580 2 x 80
1700 1640 - 1760 1900 2 x 50
2000 1930 - 2070 2230 2 x 50
2360 2280 - 2440 2610 2 x 40
2800 2710 - 2890 3090 2 x 40
3350 3240 - 3460 3670 2 x 40