|
Certification No: 181320295 |
|||
|
CERTIFICATION OF CALIBRATION |
|||
|
The Weight(s) listed below have been compared with known standards of mass.
The estimated uncertainties listed are expressed at 95% confidence level. |
|||
| TEST REF: | NOMINAL MASS g | MEASURED VALUE g | UNCERT +/- MG |
| 181320295 | 1 | 1.0000197 | 0.022 |
Signed |
|||
|
|||
| COULTER - Latex Size: 3mm (Page 1) | ||
|
COULTER CALIBRATION STANDARD |
||
| Assay Sheet | ||
| INTENDED USE: | ||
| Polymer latex particles intended for the calibration of COULTER COUNTER instruments. Other than those to be used for the red and white blood cell sizing or counting. | ||
![]() |
BATCH |
Z.12 |
| MATERIAL |
POLYSTYRENE
LATEX |
|
| DO NOT USE AFTER | 31st MARCH 1998 | |
| Other parameters: |
![]() |
| SINGLET NUMBER
MEDIAN DIAMETER (Non-multichannel instruments;Figure B). |
|
| 3.24 µm DIAMETER | |
| WEIGHT PEAK SPLIT (Models TA/TA II; Figure C). |
|
| 3.26 µm DIAMETER | |
| 1:2 WEIGHT PEAK SPLIT (Models TA/TA II; Figure D). |
|
| 3.18 µm DIAMETER | |
| NUMBER PEAK SPLIT (Models TA II/PCA (PCA 1); Figure E). |
|
| 3.22 µm DIAMETER | |
|
TYPICAL USAGE RATE: |
|
|
30µm aperture, |
2 drops per 50ml 1 drop per 100ml |
|
CALIBRATION TRACEABILITY - QUALITY ASSURANCE: |
|
|
ASSAY VALUES have been determined using COULTER COUNTER models MULTISIZER, TA II/PCA, ZM and CHANNELYZER 256 which are verified for performance and calibrated and checked with primary reference standard particles from the National Institute of Standards and Technology, N.I.S.T., (formerly National Bureau of Standards, N.B.S.), and the Community Bureau of Reference, B.C.R. (1). Property documented records of the calibrations performed with this Coulter Calibration Standard, when used as recommended in the relevant Instrument Instruction Manual, will allow traceability of those calibrations to the N.I.S.T. (N.B.S.) and B.C.R. particle size Reference Materials, as required by some National authorities, such as BS 5750 and NAMAS in the United Kingdom. CALIBRATION STANDARD SUSPENSION: SIZE CALIBRATION: Using a series of sensing apertures and latex particles. Harfield, Wharton and Lines (4) verified that the response of a COULTER COUNTER Model ZM was linear with particle volume to some 80% of the aperture diameter. By using a series of apertures and verified COULTER COUNTER models, each calibrated to the mode of the size distribution of the relevant N.I.S.T. (N.B.S.) Standard or B.C.R. Certified latex reference materials (1), this Coulter Calibration Standard has been assigned the values given above. Instruments of high definition size distribution (e.g. MULTISIZER, CHANNELYZER models) are most conveniently calibrated using the number mode size (diameter or volume), see figure A above. Single or double threshold instruments (e.g. Models A, B, D
Industrial, ZB, and ZM without CHANNELYZER accessory) are more conveniently
calibrated using the singlet number median diameter, see Figure B above.
The 'weight peak split' method (Figure C) is intended for
calibration of TA and TAII Models without a population Count Accessory, whereby
the calibration point is the junction of two consecutive channels containing
equal weight (volume). Earlier calibration methods for the TA or TAII used the 1:2
weight peak split method (Figure D) whereby the calibration point is the
junction of two consecutive channels containing twice the weight (volume) in the
right hand channel as in the left. The 'number peak split' method (Figure E) is intended for the
Model TAII with PCA or PCA 1 population count accessory. |
|
|
|
|
|
CHOICE OF CALIBRATION STANDARD: REDISPERSION OF CALIBRATION STANDARDS: Dry calibration standards must be dispersed by first spatulating with a surfactant, such as Coulter Dispersant. Further dispersion with ultrasonic energy is recommended. FREQUENCY OF CALIBRATION: INDICATION OF PRODUCT INSTABILITY OR DETERIORATION: (a) Check that the aperture is clear, and that the wafer and
the aperture tube are clean and free from bubbles. HEALTH AND SAFETY: REFERENCES: WARRANTY: |
|
| Coulter Electronics Limited, Northwell Drive, Luton, Beds., LU3 3RH England Tel: (0582) 491414 Fax: (0582) 390390 |
![]() |
|
|
|
| COULTER - Latex Size: 5mm (Page 2) | ||
|
COULTER CALIBRATION STANDARD |
||
| Assay Sheet INTENDED USE: |
||
|
Polymer latex particles intended for the calibration of COULTER COUNTER instruments. Other than those to be used for the red and white blood cell sizing or counting. |
||
![]() |
BATCH |
C.19 |
| MATERIAL | P.D.V.B.
LATEX |
|
| DO NOT USE AFTER | 30th June 1998 | |
| Other parameters: |
![]() |
|
|
|
|
SINGLET NUMBER
MEDIAN DIAMTER |
|
|
5.24 µm DIAMETER |
|
|
|
|
|
WEIGHT PEAK
SPLIT |
|
|
5.22 µm DIAMETER |
|
|
|
|
|
1:2 WEIGHT
PEAK SPLIT |
|
|
5.12 µm DIAMETER |
|
|
|
|
|
NUMBER PEAK
SPLIT |
|
|
5.19 µm DIAMETER |
|
|
TYPICAL USAGE RATE: |
|
|
70µm aperture, |
3 drops per 50ml |
|
100µm aperture, |
1 drop per 50ml |
|
CALIBRATION TRACEABILITY - QUALITY ASSURANCE: Property documented records of the calibrations performed with this Coulter Calibration Standard, when used as recommended in the relevant Instrument Instruction Manual, will allow traceability of those calibrations to the N.I.S.T. (N.B.S.) and B.C.R. particle size Reference Materials, as required by some National authorities, such as BS 5750 and NAMAS in the United Kingdom. CALIBRATION STANDARD SUSPENSION: SIZE CALIBRATION: Using a series of sensing apertures and latex particles. Harfield, Wharton and Lines (4) verified that the response of a COULTER COUNTER Model ZM was linear with particle volume to some 80% of the aperture diameter. By using a series of apertures and verified COULTER COUNTER models, each calibrated to the mode of the size distribution of the relevant N.I.S.T. (N.B.S.) Standard or B.C.R. Certified latex reference materials (1), this Coulter Calibration Standard has been assigned the values given above. Instruments of high definition size distribution (e.g. MULTISIZER, CHANNELYZER models) are most conveniently calibrated using the number mode size (diameter or volume), see figure A above. Single or double threshold instruments (e.g. Models A, B, D Industrial, ZB, and ZM without CHANNELYZER accessory) are more conveniently calibrated using the singlet number median diameter, see Figure B above. The 'weight peak split' method (Figure C) is intended for calibration of TA and TAII Models without a population Count Accessory, whereby the calibration point is the junction of two consecutive channels containing equal weight (volume). Earlier calibration methods for the TA or TAII used the 1:2 weight peak split method (Figure D) whereby the calibration point is the junction of two consecutive channels containing twice the weight (volume) in the right hand channel as in the left. The 'number peak split' method (Figure E) is intended for the Model TAII with PCA or PCA 1 population count accessory. |
|
|
|
CHOICE OF CALIBRATION STANDARD: REDISPERSION OF CALIBRATION STANDARDS: Dry calibration standards must be dispersed by first spatulating with a surfactant, such as Coulter Dispersant. Further dispersion with ultrasonic energy is recommended. FREQUENCY OF CALIBRATION: INDICATION OF PRODUCT INSTABILITY OR DETERIORATION: (a) Check that the aperture is clear, and that the wafer and
the aperture tube are clean and free from bubbles. HEALTH AND SAFETY: REFERENCES: |
|
| Coulter
Electronics Limited, Northwell Drive, Luton, Beds., LU3 3RH England Tel: (0582) 491414 Fax: (0582) 390390 |
![]() |
| COULTER - Latex Size 10mm (Page 3) | ||
|
COULTER CALIBRATION STANDARD |
||
| Assay Sheet INTENDED USE: |
||
|
Polymer latex particles intended for the calibration of COULTER COUNTER instruments. Other than those to be used for the red and white blood cell sizing or counting. |
||
![]() |
BATCH | D.42 |
| MATERIAL | P.D.V.B. LATEX | |
| DO NOT USE AFTER | 31st January 1998 | |
| Other parameters: |
![]() |
|
SINGLET NUMBER
MEDIAN DIAMTER |
|
|
10.0 µm DIAMETER |
|
|
WEIGHT PEAK
SPLIT |
|
|
10.1 µm DIAMETER |
|
|
1:2 WEIGHT
PEAK SPLIT |
|
|
9.8 µm DIAMETER |
|
|
NUMBER PEAK
SPLIT |
|
|
10.0 µm DIAMETER |
|
|
TYPICAL USAGE RATE: |
|
|
100µm aperture, |
1-2 drops per 100ml |
|
140µm aperture, |
1 drop per 150ml |
|
CALIBRATION TRACEABILITY - QUALITY ASSURANCE: Property documented records of the calibrations performed with this Coulter Calibration Standard, when used as recommended in the relevant Instrument Instruction Manual, will allow traceability of those calibrations to the N.I.S.T. (N.B.S.) and B.C.R. particle size Reference Materials, as required by some National authorities, such as BS 5750 and NAMAS in the United Kingdom. CALIBRATION STANDARD SUSPENSION: SIZE CALIBRATION: Using a series of sensing apertures and latex particles.
Harfield, Wharton and Lines (4) verified that the response of a COULTER COUNTER
Model ZM was linear with particle volume to some 80% of the aperture diameter.
By using a series of apertures and verified COULTER COUNTER models, each
calibrated to the mode of the size distribution of the relevant N.I.S.T. (N.B.S.)
Standard or B.C.R. Certified latex reference materials (1), this Coulter
Calibration Standard has been assigned the values given above. Instruments of high definition size distribution (e.g.
MULTISIZER, CHANNELYZER models) are most conveniently calibrated using the
number mode size (diameter or volume), see figure A above. Single or double threshold instruments (e.g. Models A, B, D
Industrial, ZB, and ZM without CHANNELYZER accessory) are more conveniently
calibrated using the singlet number median diameter, see Figure B above. |
|
|
|
REDISPERSION OF CALIBRATION STANDARDS: FREQUENCY OF CALIBRATION: INDICATION OF PRODUCT INSTABILITY OR DETERIORATION: (a) Check that the aperture is clear, and that the wafer and
the aperture tube are clean and free from bubbles. HEALTH AND SAFETY: REFERENCES: 1. (a) SRM 1960, nominal 10µm latex;
SRM 1961, Nominal 30µm latex; WARRANTY: |
|
| Coulter
Electronics Limited, Northwell Drive, Luton, Beds., LU3 3RH England Tel: (0582) 491414 Fax: (0582) 390390 |
|
| COULTER - Latex Size: 20mm (Page 4) | ||
|
COULTER CALIBRATION STANDARD |
||
| Assay Sheet INTENDED USE: |
||
| Polymer latex particles intended for the calibration of COULTER COUNTER instruments. Other than those to be used for the red and white blood cell sizing or counting. | ||
![]() |
BATCH | F.32 |
|
MATERIAL |
P.D.V.B. LATEX | |
| DO NOT USE AFTER | 30st September 1998 | |
| Other parameters: |
![]() |
| SINGLET NUMBER
MEDIAN DIAMTER (Non-multichannel instruments; Figure B). |
|
| 21.4 µm DIAMETER | |
| WEIGHT PEAK
SPLIT (Models TA/TA II; Figure C). |
|
| 22.3 µm DIAMETER | |
| 1:2 WEIGHT
PEAK SPLIT (Models TA/TA II; Figure D). |
|
| 21.3 µm DIAMETER | |
| NUMBER PEAK
SPLIT (Models TA II/PCA (PCA 1); Figure E). |
|
| 21.7 µm DIAMETER | |
|
TYPICAL USAGE RATE: |
|
| 200µm aperture, | 1-2 drops per 100ml |
| 280µm aperture, | 1 drop per 100ml |
|
CALIBRATION TRACEABILITY - QUALITY ASSURANCE: Property documented records of the calibrations performed with this Coulter Calibration Standard, when used as recommended in the relevant Instrument Instruction Manual, will allow traceability of those calibrations to the N.I.S.T. (N.B.S.) and B.C.R. particle size Reference Materials, as required by some National authorities, such as BS 5750 and NAMAS in the United Kingdom. CALIBRATION STANDARD SUSPENSION: SIZE CALIBRATION: In practice, it is more convenient to calibrate using narrow range particles which have already been sized by other techniques. The most common method is to calibrate with spherical polymer latex particles; Secondary Calibration procedure of BS 3406: Part: 1983 (3). Using a series of sensing apertures and latex particles. Harfield, Wharton and Lines (4) verified that the response of a COULTER COUNTER Model ZM was linear with particle volume to some 80% of the aperture diameter. By using a series of apertures and verified COULTER COUNTER models, each calibrated to the mode of the size distribution of the relevant N.I.S.T. (N.B.S.) Standard or B.C.R. Certified latex reference materials (1), this Coulter Calibration Standard has been assigned the values given above. Instruments of high definition size distribution (e.g. MULTISIZER, CHANNELYZER models) are most conveniently calibrated using the number mode size (diameter or volume), see figure A above. Single or double threshold instruments (e.g. Models A, B, D Industrial, ZB, and ZM without CHANNELYZER accessory) are more conveniently calibrated using the singlet number median diameter, see Figure B above. The 'weight peak split' method (Figure C) is intended for calibration of TA and TAII Models without a population Count Accessory, whereby the calibration point is the junction of two consecutive channels containing equal weight (volume). Earlier calibration methods for the TA or TAII used the 1:2 weight peak split method (Figure D) whereby the calibration point is the junction of two consecutive channels containing twice the weight (volume) in the right hand channel as in the left. The 'number peak split' method (Figure E) is intended for the Model TAII with PCA or PCA 1 population count accessory. |
|
|
|
CHOICE OF CALIBRATION STANDARD: REDISPERSION OF CALIBRATION STANDARDS: FREQUENCY OF CALIBRATION: INDICATION OF PRODUCT INSTABILITY OR DETERIORATION: (a) Check that the aperture is clear, and that the wafer and
the aperture tube are clean and free from bubbles. HEALTH AND SAFETY: REFERENCES: |
|
| Coulter
Electronics Limited, Northwell Drive, Luton, Beds., LU3 3RH England Tel: (0582) 491414 Fax: (0582) 390390 |
![]() |
| COULTER - Latex Size 40mm (Page 5) | ||
|
COULTER CALIBRATION STANDARD |
||
| Assay Sheet INTENDED USE: |
||
|
Polymer latex particles intended for the calibration of COULTER COUNTER instruments. Other than those to be used for the red and white blood cell sizing or counting. |
||
|
|
BATCH |
G.18 |
| MATERIAL | P.D.V.B. LATEX | |
| DO NOT USE AFTER | ||
| Other parameters: |
![]() |
| SINGLET NUMBER
MEDIAN DIAMTER (Non-multichannel instruments; Figure B). |
|
| 41.2 µm DIAMETER | |
| WEIGHT PEAK
SPLIT (Models TA/TA II; Figure C). |
|
| 43.3 µm DIAMETER | |
| 1:2 WEIGHT
PEAK SPLIT (Models TA/TA II; Figure D). |
|
| 41.0 µm DIAMETER | |
| NUMBER PEAK
SPLIT (Models TA II/PCA (PCA 1); Figure E). |
|
| 41.9 µm DIAMETER | |
|
TYPICAL USAGE RATE: |
|
| 400µm aperture, | 1-2 drops per 100ml |
| 560µm aperture, | 1 drop per 200ml |
|
CALIBRATION TRACEABILITY - QUALITY ASSURANCE: CALIBRATION STANDARD SUSPENSION: SIZE CALIBRATION: Using a series of sensing apertures and latex particles. Harfield, Wharton and Lines (4) verified that the response of a COULTER COUNTER Model ZM was linear with particle volume to some 80% of the aperture diameter. By using a series of apertures and verified COULTER COUNTER models, each calibrated to the mode of the size distribution of the relevant N.I.S.T. (N.B.S.) Standard or B.C.R. Certified latex reference materials (1), this Coulter Calibration Standard has been assigned the values given above. Instruments of high definition size distribution (e.g. MULTISIZER, CHANNELYZER models) are most conveniently calibrated using the number mode size (diameter or volume), see figure A above. Single or double threshold instruments (e.g. Models A, B, D Industrial, ZB, and ZM without CHANNELYZER accessory) are more conveniently calibrated using the singlet number median diameter, see Figure B above. The 'weight peak split' method (Figure C) is intended for
calibration of TA and TAII Models without a population Count Accessory, whereby
the calibration point is the junction of two consecutive channels containing
equal weight (volume). Earlier calibration methods for the TA or TAII used the 1:2
weight peak split method (Figure D) whereby the calibration point is the
junction of two consecutive channels containing twice the weight (volume) in the
right hand channel as in the left. The 'number peak split' method (Figure E) is intended for the
Model TAII with PCA or PCA 1 population count accessory. |
|
|
|
|
REDISPERSION OF CALIBRATION STANDARDS: FREQUENCY OF CALIBRATION: INDICATION OF PRODUCT INSTABILITY OR DETERIORATION: (a) Check that the aperture is clear, and that the wafer and
the aperture tube are clean and free from bubbles. HEALTH AND SAFETY: REFERENCES: WARRANTY: |
|
| Coulter
Electronics Limited, Northwell Drive, Luton, Beds., LU3 3RH England Tel: (0582) 491414 Fax: (0582) 390390 |
![]() |
|
UNITED STATES DEPARTMENT OF COMMERCE |
|
|
REPORT OF CALIBRATION |
|
|
For: Item: |
Whitehouse Scientific Ltd. Whitchurch Road, Waverton Chester, CH3 7PB, England 10 MM Stage Micrometer Graticules, Ltd., NIST #5594 |
|
This stage micrometer was calibrated with the NIST Line Scale Interferometer (LSI). The LSI consists of a scanning electro-optical line detector, a high precision one-axis motion system, and a high accuracy heterodyne interferometer for determining the displacement of the test artifact beneath the line detector. The wavelength of a stabilized helium-neon laser corrected for temperature, humidity and atmospheric pressure is used as the length standard. The instrument is housed in an environmental chamber in which all environmental properties are carefully monitored. The complete description of the design and operation of the NIST LSI is given in the Journal of Research of the National Institute of Standards and Technology Volume 104, Number 3, May-June 1999, "The NIST Length Scale Interferometer". Results of the calibration are given on the following pages of this report. Each length value is the mean of 4 measurements and the expanded uncertainty in each value is |
|
|
U=kuc |
|
| Where the combined standard uncertainty | |
|
uc = Ö (ui2 +uj2) |
|
|
Where ui is the standard uncertainty arising from random effects and uj is the standard uncertainty arising from systematic effects in the measurement process. The coverage factor k=2 was used which gives for each reported value a level of confidence of approximately 95 percent. Measurements were made from line center to line center using a 0.16 mm segment of each. |
|
|
NIST Test No. 821/263573-00 |
|
Graduation line midway between the line tips. During measurement the stage micrometer was placed on a flat surface. During the measurement the environmental chamber and the artifact temperature
were held within +/-0.005°C of 20°Celsius. All lengths are reported at a
temperature of 20°Celsius (68°Fahrenheit). A coefficient of linear thermal
expansion of 8.5 x 10-6 / °Celsius was used in normalizing the lengths to
20°Celsius. |
Certified correct by |
For the Director |
|
NIST Test No. 821/263573-00 |
|||||
|
INTERVAL |
LENGTH |
CORRECTION |
ui |
uj |
U |
|
0.000 TO 0.100 |
0.099837 |
-0.000163 |
0.000002 |
0.000000 |
0.000004 |
|
0.000 TO 0.200 |
0.199853 |
-0.000147 |
0.000003 |
0.000000 |
0.000006 |
|
0.000 TO 0.300 |
0.299811 |
-0.000189 |
0.000003 |
0.000000 |
0.000006 |
|
0.000 TO 0.400 |
0.399828 |
-0.000172 |
0.000003 |
0.000000 |
0.000007 |
|
0.000 TO 0.500 |
0.499904 |
-0.000096 |
0.000005 |
0.000000 |
0.000009 |
|
0.000 TO 0.600 |
0.600004 |
0.000004 |
0.000005 |
0.000000 |
0.000011 |
|
0.000 TO 0.700 |
0.700009 |
0.000009 |
0.000005 |
0.000000 |
0.000010 |
|
0.000 TO 0.800 |
0.800053 |
0.000053 |
0.000003 |
0.000000 |
0.000006 |
|
0.000 TO 0.900 |
0.900036 |
0.000036 |
0.000003 |
0.000000 |
0.000006 |
|
0.000 TO 1.000 |
1.000075 |
0.000075 |
0.000002 |
0.000000 |
0.000005 |
|
|||||
|
NIST Test No. 821/263573-00 |
|||||
| INTERVAL (MILLIMETER) |
LENGTH (mm) |
CORRECTION (mm) |
ui (mm) |
uj (mm) |
U (mm) |
| 0.000 TO 1.00 | 1.000080 |
0.000080 |
0.000002 | 0.000000 | 0.000004 |
| 0.000 TO 2.00 | 2.000233 | 0.000233 | 0.000008 | 0.000000 | 0.000015 |
| 0.000 TO 3.00 | 3.000448 | 0.000448 | 0.000006 | 0.000000 | 0.000012 |
| 0.000 TO 4.00 | 4.000759 | 0.000759 | 0.000006 | 0.000000 | 0.000012 |
| 0.000 TO 5.00 | 5.001022 | 0.001022 | 0.000006 | 0.000000 | 0.000013 |
| 0.000 TO 6.00 | 6.001336 | 0.001336 | 0.000004 | 0.000000 | 0.000009 |
| 0.000 TO 7.00 | 7.001562 | 0.001562 | 0.000005 | 0.000000 | 0.000010 |
| 0.000 TO 8.00 | 8.001715 | 0.001715 | 0.000001 | 0.000000 | 0.000002 |
| 0.000 TO 9.00 | 9.001707 | 0.001707 | 0.000003 | 0.000000 | 0.000006 |
| 0.000 TO 10.00 | 10.001704 | 0.001704 | 0.000003 | 0.000001 | 0.000005 |
|
|||||
NPL CENTRE FOR MECHANICAL AND OPTICAL TECHNOLOGY Certificate of Calibration |
|
|
REFERENCE STAGE GRATICULE |
|
FOR |
Whitehouse Scientific Ltd Whitchurch Road Waverton Chester CH3 7PB For the attention of Dr Graham Rideal, |
|
DESCRIPTION 1) A square grid comprising boxes varying in size from 25 µm to 400 µm. The
numbers in the centre of the squares represent the nominal length of the sides
of the square in which they appear. The uncertainties are for a confidence probability of not less than 95%. DATE OF CALIBRATION
18 June 1997 |
|
| Date of issue: 18 June 1997 Signed: (Authorised Signatory) |
![]() |
|
Checked by: |
|
Name: Nicholas P Turner For Managing Director |
|
This certificate provides traceability of measurement to recognised national standards, and to the units of measurement realised at the NPL or other recognised national standards laboratories. This certificate may only be published in full, unless permission for publication of an approved extract has been obtained in writing from the Managing Director. It does not itself impute to the subject of calibration any attributes beyond those shown by the data contained herein. |
|
NATIONAL PHYSICAL LABORATORY |
|||
|
MEASUREMENTS The spots were measured using a microscope with a 100x objective and an image analyser. They were compared with spots measured using an image-shearing microscope, calibrated against an interferometrically measured stage micrometer. The diameters are calculated from the area of the spots and thus represent an average spot diameter. RESULTS: |
|||
| Spot Number | Diameter (µm) | ||
| 1 | 2.53 | ||
| 2 | 3.75 | ||
| 3 | 5.57 | ||
| 4 | 8.09 | ||
| 5 | 11.63 | ||
| 6 | 16.61 | ||
| 7 | 23.64 | ||
| 8 | 33.5 | ||
| 9 | 47.67 | ||
| 4) The log-normal distribution has a mean of 10.58 µm and a standard
deviation of 1.51 µm. |
|||
| Reference: 08A038/970127/106-66 | |||
| Checked by: |
|||
|
NATIONAL PHYSICAL LABORATORY |
|
UNCERTAINTIES: NOTES: |
| Reference: 08A038/970127/106-66 |
| Checked by: |
|
NPL |
|
|
USE OF THE NPL REFERENCE STAGE GRATICULE |
|
| The reference stage graticule comprises four test areas. They are as follows: | |
1.
THE GRID |
|
|
|
2.
THE MONOSIZE ARRAY |
|
the spots at the edges can be seen by comparison with spots in the middle of the screen. Measurements must be made rather than direct visual observations because screen distortions may affect judgement. |
|
|
3. THE ROOT-2 ARRAY |
|
|
|
The root-2 array is used to determine the threshold level required to measure the spots correctly. The image analyser works by the user choosing a grey level at which anything darker is spot and anything lighter is |
|
background. As the edge of a spot is not a clean edge but a blurred one due to the limited resolution of the camera and the physical limits of optical microscopy, the specified threshold level affects the measured size of the spots. This array of spots provides a useful research tool for investigating the effects of background lighting, detect level, focus etc. on measured spot size. |
|
|
|
4.
THE LOG-NORMAL ARRAY |
|
The equations that define the mean and standard deviation of the log-normal
distribution are: |
|
| ( | 100 | log | ) |
| å | ___ | ||
| i-1 | 100 |
|
STANDARD DEVIATION = exp |
|
( |
100 | (log MEAN-logdi )2 | )1/2 | ||
| å | ______________ | ||||
| i-1 |
100 |
||||
|
Where di is the diameter of spot number I. |
|
Please contact NPL if you wish the log-normal parameters to be calculated using alternative formulae. |
| Always ensure the graticule is clean before use. |
|
CENTRE FOR MECHANICAL AND OPTICAL TECHNOLOGY |